Settings
............................
5.5.3 – 5.5.3.1
5.5.3 Measurement parameters for dot gain
........................................................................................................................................
One to three halftone reference values in the range between 0% and 100% can be entered. At
the upper and lower ends of this range, the selected halftone values can be switched off. This
will be confirmed with <off> on the display.
........................................................................................................................................
5.5.3.1 Entry of halftone reference values
Procedure:
Select the menu line Dot gain by turning the <Control ball>.
Activate the selected function by pressing the <Measurement key>.
An entry field for three halftone percentages is shown on the display.
Select the menu line 1.Screen percent and confirm the selection by pressing the <Measurement
key>.
The entry field for the 1st halftone reference value is shown on the display.
Enter the 1st halftone reference value by turning the <Control ball> and confirm the entry by
pressing the <Measurement key>.
Select the menu line 2.Screen percent and confirm the selection by pressing the <Measurement
key>.
The entry field for the 2nd halftone reference value is shown on the display.
Enter the 2nd halftone reference value by turning the <Control ball> and confirm the entry by
pressing the <Measurement key>.