Device settings
............................
6.2 – 6.3
6.2 Turn display
........................................................................................................................................
The display of the D190 remission densitometer can be reversed 180°.
In this way, the unit can be operated with either the left or the right hand.
........................................................................................................................................
Procedure:
Select the menu line Turn display by turning the <Control ball>.
Activate the selected function by pressing the <Measurement key>.
The display on the screen is reversed 180 °.
Select the menu line Return by turning the <Control ball> and press the <Measurement key>
several times to return to the main menu.
6.3 LCD contrast
........................................................................................................................................
On the D190 remission densitometer, the contrast of the display can be adjusted.
........................................................................................................................................
Procedure:
Select the menu line LCD contrast by turning the <Control ball>.
Activate the selected function by pressing the <Measurement key>.
The display shows the entry field for the contrast adjustment.
Turn the <Control ball> to obtain the optimum contrast setting and terminate the entry by pressing
the <Measurement key>.
Select the menu line Return by turning the <Control ball> and press the <Measurement key>
several times to return to the main menu.