Densitometric functions
............................
4.5.1
Select the menu line Measure paper white by turning the <Control ball>.
Position the measuring aperture on the paper white and initiate a measurement by pressing the
<Measurement key>.
The display White measurement done is shown for about 1 second.
Select the menu line Measure ‘color’ by turning the <Control ball>.
Position the measuring aperture on the color sample and initiate the measurements by pressing the
<Measurement key>.
The density values of the measured color are shown on the display.
Select the menu line where you want to change the name ore the value of the reference.
Activate the selection by pressing the <Measurement key>.
Change the values.
Activate the changes by pressing the <Measurement key>.
Repeat all this steps until all values of the reference are corrected.
Repeat all this steps until all colors are defined as references.