Device settings
............................
6.11 – 6.11.1
6.11 Electronic safeguard
........................................................................................................................................
The D190 remission densitometer is equipped with an electronic safeguard. It prevents the
measuring module from being extended if vibrations occur during transport.
........................................................................................................................................
........................................................................................................................................
The Electronic safeguard can be switched on/off under Settings k Device settings k
Transport protection.
........................................................................................................................................
6.11.1 Switching on the electronic safeguard
Procedure:
Select the menu line Settings by turning the <Control ball> and confirm the selection by pressing
the <Measurement key>.
Select the menu line Device settings by turning the <Control ball> and confirm the selection by
pressing the <Measurement key>.
Select the menu line Transport protection by turning the <Control ball>.
Switch on the electronic safeguard by pressing the <Measurement key>.
The display shows a code number for the release of the electronic safeguard.